Applications and Methods:
-
Diagnostics only
-
Diagnostics and Asynchronous Esophageal pacing (no extra stimuli)
-
Diagnostics and Asynchronous Esophageal pacing with multiple extra stimuli,
Diagnostics only
The bipolar esophageal ECG is an excellent complement to the std 12 lead ECG recording.
Typical applications: Acute and/or scheduled diagnostics of atrial arrhythmias.
Equipment required: Bipolar esophageal lead, Rostock Filter and standard ECG equipment or any standard Monitoring system.
Diagnostics and Esophageal pacing
Requirements / necessary equipment & functions:
Signal acquisition
High Pass filter with choice of multiple settings of the frequency cut off and variable amplification of the signal.
Esophageal pacing
Asynchronous external stimulator without or with the extra stimuli facility.
Constant current source, pulse width 10 ms and output range of 5 – 25 mA.
Diagnostics and Asynchronous Esophageal pacing (no extra stimuli)
Bipolar ESOPHAGEAL REGISTRATION and ESOPHAGEAL PACING
Typical applications: Acute and/or scheduled diagnostics and treatment of atrial arrhythmias
such as overdrive pace of Atrial Flutter
Equipment required: ESOPHAGEAL leads/4-polar, Rostock Filter, Asynchronous esophageal stimulator and std ECG equipment
Diagnostics and Asynchronous Esophageal pacing with multiple extra stimuli,
Bipolar ESOPHAGEAL REGISTRATION and ESOPHAGEAL PACING
Typical applications: Acute and/or scheduled diagnostics, treatment of atrial arrhythmias
(overdrive pace of Atrial Flutter) and complete semi-invasive investigations according to protocol with extra stimuli
Equipment required: ESOPHAGEAL leads/4-polar, Rostock Filter, Asynchronous esophageal stimulator with multiple extra stimuli and std ECG equipment